Yamato Laboratory Incubator | IJ101
Features of Low Temperature Incubator (Programmable, Peltier Cooling) (IJ101)
Specifications
Item Code | 221164 |
---|---|
Model | IJ101 |
Type | Forced convection |
Operating Temperature Range | 5~60?(at room temp.25?) |
Temp. Control Accuracy | ±0.3?(at 37?) |
Temp. Distribution Accuracy | ±1.0?(at 37?) |
Time to Max. Temp. | 20~60? 60min. |
Time to Lowest Temp. | 20~5? 100min. |
Interior | Stainless steel |
Heater | Mica heater 120W |
Condenser | Peltiert element, Forced radiation type |
Temp. Control Method | PID |
Operation Function | Program operation (30 steps x 1, 15 steps x 2, 10 steps x 3), Fixed operation, Quick auto-stop, Auto-stop, Auto-start operation |
Additional Functions | Calibration off-set, Key-lock function, Auto recovery function at power failure |
Sensor | Double sensor: Platinum resistance Pt 100Ω (Temp. controller), K-thermocouple(Overheat preventor) |
Safety Functions | Self-diagnostic function(Temp. sensor error, heater failure, SSR short-circuit, Auto overheat), Overheat prevention(IJ201/301), Liquid pressure overheat preventor, Over-current ELB, Main relay error |
Shelf Loading Capacity | Approx. 5kg/ shelf |
Shelf Step | 7 steps |
Accessories | Shelf(Stainless punching metal) 2pcs., Shelf support 4 pcs., Receiver for drain 1pc. |
Internal dimensions(W×D×H mm) | 250×250×250mm |
External dimensions(W×D×H mm) | 350×399×565mm |
Remarks(dimensions) | 15.6L |
Power Source | AC115V/AC220V single phase with step-down transformer |
Weight | 20kg |
Maintenance and Warranty
Conditions: temperature and humidity: 23 deg.C+/-5 deg.C, 65%RH+/-20% (no load)
External dimensions do not include protrusions.
To contact Banksia Scientific Company about Yamato Laboratory Incubator | IJ101 use Get a quote.
Features of Low Temperature Incubator (Programmable, Peltier Cooling) (IJ101)
Specifications
Item Code | 221164 |
---|---|
Model | IJ101 |
Type | Forced convection |
Operating Temperature Range | 5~60?(at room temp.25?) |
Temp. Control Accuracy | ±0.3?(at 37?) |
Temp. Distribution Accuracy | ±1.0?(at 37?) |
Time to Max. Temp. | 20~60? 60min. |
Time to Lowest Temp. | 20~5? 100min. |
Interior | Stainless steel |
Heater | Mica heater 120W |
Condenser | Peltiert element, Forced radiation type |
Temp. Control Method | PID |
Operation Function | Program operation (30 steps x 1, 15 steps x 2, 10 steps x 3), Fixed operation, Quick auto-stop, Auto-stop, Auto-start operation |
Additional Functions | Calibration off-set, Key-lock function, Auto recovery function at power failure |
Sensor | Double sensor: Platinum resistance Pt 100Ω (Temp. controller), K-thermocouple(Overheat preventor) |
Safety Functions | Self-diagnostic function(Temp. sensor error, heater failure, SSR short-circuit, Auto overheat), Overheat prevention(IJ201/301), Liquid pressure overheat preventor, Over-current ELB, Main relay error |
Shelf Loading Capacity | Approx. 5kg/ shelf |
Shelf Step | 7 steps |
Accessories | Shelf(Stainless punching metal) 2pcs., Shelf support 4 pcs., Receiver for drain 1pc. |
Internal dimensions(W×D×H mm) | 250×250×250mm |
External dimensions(W×D×H mm) | 350×399×565mm |
Remarks(dimensions) | 15.6L |
Power Source | AC115V/AC220V single phase with step-down transformer |
Weight | 20kg |
Maintenance and Warranty
Conditions: temperature and humidity: 23 deg.C+/-5 deg.C, 65%RH+/-20% (no load)