Tecnair Proportional Regulator Electropneumatic ...
The lower range model is designed for standard process control applications which typically utilize a 3 to 15 psig output. The extended range unit provides up to 120 psig output for higher pressure industrial pneumatic and process control system requirements.
The transducers convert a current or voltage input signal to a linearly proportional pneumatic output pressure. This versatile instrument is designed for control applications that require a high degree of reliability and repeatability at an economical cost. Optional NEMA 4X (IP65) version allows for splashdown and outdoor installation. Typically, these units are used for applications that require the operation of valve actuators, pneumatic valve positioners, damper and louver actuators, final control elements, relays, air cylinders, web tensioners, clutches, and brakes. Industries that utilize the Type 500X include Petrochemical, HVAC, Energy Management, Textile, Paper, Paper Converting, Food and Drug.
TECHNICAL DATA | |||
---|---|---|---|
Model |
Low pressure series |
High pressure series |
|
Port Size | G1/4 | G1/4 | |
Pressure Supply | 7 bar max | 10.5 bar max | |
Terminal Bases Linearity | <±0.75% of Corresponding Range | <±1.5% of Corresponding Range | |
Repeatability | <±0.5% of Corresponding Range | <±0.5% of Corresponding Range | |
Hysteresis | <±1.0% of Corresponding Range | <±0.5% of Corresponding Range | |
Flow Rate | 7.6 m3/hr (supply: 1.7bar) | 34 m3/hr (supply: 10.5bar) | |
20 m3/hr (supply: 6.8bar) | |||
Relief Capacity | 3.4 m3/hr at 0.35bar | 11.9 m3/hr at 0.7bar | |
(above 1.3bar setpoint) | (above 1.3bar setpoint) | ||
Air Consumption (max) | 0.07 m3/hr (midrange) | 14 m3/hr (midrange) | |
Weight | 0.94kg | 0.94kg | |
Medium | 40 Micron Instrument Compressed Air (oil free) & Neutral Gases | ||
Operating Temperature | -30° > +60°C | ||
Electrical Connection | Cable Gland G1/2 (various connectors available upon request) | ||
Filtration | 50 μm Minimum Filtration | ||
IP Rating | IP 65 upon request (NEMA-4X enclosure required) |
To contact Tecnair about Tecnair Proportional Regulator Electropneumatic Transducer | PR500X Series use Get a quote.
The lower range model is designed for standard process control applications which typically utilize a 3 to 15 psig output. The extended range unit provides up to 120 psig output for higher pressure industrial pneumatic and process control system requirements.
The transducers convert a current or voltage input signal to a linearly proportional pneumatic output pressure. This versatile instrument is designed for control applications that require a high degree of reliability and repeatability at an economical cost. Optional NEMA 4X (IP65) version allows for splashdown and outdoor installation. Typically, these units are used for applications that require the operation of valve actuators, pneumatic valve positioners, damper and louver actuators, final control elements, relays, air cylinders, web tensioners, clutches, and brakes. Industries that utilize the Type 500X include Petrochemical, HVAC, Energy Management, Textile, Paper, Paper Converting, Food and Drug.
TECHNICAL DATA | |||
---|---|---|---|
Model |
Low pressure series |
High pressure series |
|
Port Size | G1/4 | G1/4 | |
Pressure Supply | 7 bar max | 10.5 bar max | |
Terminal Bases Linearity | <±0.75% of Corresponding Range | <±1.5% of Corresponding Range | |
Repeatability | <±0.5% of Corresponding Range | <±0.5% of Corresponding Range | |
Hysteresis | <±1.0% of Corresponding Range | <±0.5% of Corresponding Range | |
Flow Rate | 7.6 m3/hr (supply: 1.7bar) | 34 m3/hr (supply: 10.5bar) | |
20 m3/hr (supply: 6.8bar) | |||
Relief Capacity | 3.4 m3/hr at 0.35bar | 11.9 m3/hr at 0.7bar | |
(above 1.3bar setpoint) | (above 1.3bar setpoint) | ||
Air Consumption (max) | 0.07 m3/hr (midrange) | 14 m3/hr (midrange) | |
Weight | 0.94kg | 0.94kg | |
Medium | 40 Micron Instrument Compressed Air (oil free) & Neutral Gases | ||
Operating Temperature | -30° > +60°C | ||
Electrical Connection | Cable Gland G1/2 (various connectors available upon request) | ||
Filtration | 50 μm Minimum Filtration | ||
IP Rating | IP 65 upon request (NEMA-4X enclosure required) |
To contact Tecnair about Tecnair Proportional Regulator Electropneumatic Transducer | PR500X Series use Get a quote.